Laboratory

The laboratory equipment of the HiPE-LAB at the University of Bremen, consisting of the loading facility, the preliminary test platform and climate climate chambers. Extensive measurement technology and sensors are available for measuring voltages, currents, temperatures, relative humidity and air flow. For further information please contact us.

3D view on the HiPE-LAB.
? University of Bremen - Dr.-Ing. W. Holzke

Load system and preliminary test platform

The load system of the HiPE-LAB.
? University of Bremen

Technical specifications - Load system

Power range10,8 MVA  (4x 2,7 MVA)
Voltage range0 to 1.000 V AC
Current rangeup to 9.000 A AC (4 x 2.250 A AC)
Frequency range0 to 70 Hz, when operated with transformers  45 to 70 Hz

Electrical system

IT system  (isolated against earth, other electrical systems on request)
The preliminary test platform of the HiPE-LAB.
? University of Bremen

Technical specification - Preliminary test plattform

Power range300 kVA
Funktional rangeIdentical to the load system
ApplicationModel development and low power testing

Climate chambers

The open climate chamber of the HiPE-LAB.
? University of Bremen

Technical specification - Climate room

Temperature range-40 °C to +120 °C
Relative humidity range10% to 95% relative humidity at +10 °C to +95 °C
Testing room (l x b x h) 7,5 m x 5,3 m x 4,3 m
Cooling water connection for device under test (DUT)300 kW, supply temperature +16 °C to +60 °C

 

[Translate to English:] Sicht auf einen ge?ffneten Klimaschrank.
? University of Bremen

Technical specifications - Climate cabinets

Temperature range

-70 °C to +180 °C
Relative humidity range10 % to 98 % relative humidity
Testing roomup to 1000 l